Sep 06, 2010  Hi, I'm looking for the user manual of Kenwood CS-6020 Oscilloscope. I did google but had no luck. Advertisement 30th July 2007, 15:39 #2. View Forum Posts. Need service manual of Kenwood CS-4025 oscilloscope (1) Part and Inventory Search. Top Helped /. Download KENWOOD CS-1575D OSCILLOSCOPE service manual & repair info for electronics experts. Service manuals, schematics, eproms for electrical technicians. User manual for shark vac then steam. Similar manuals. KENWOOD TRIO CS-2100A KENWOOD TRIO CS-4025 OSCILOSCOPE SCH. If you want to join us and get repairing help please sign in or sign up by completing a simple electrical.

  1. Oscilloscope Instruction Manual
  2. Kenwood Cs-4025 Oscilloscope Manual
  3. Dso138 Oscilloscope Manual
  4. Kenwood Cs-4025 Oscilloscope Service Manual
  5. Oscilloscope Manual Pdf

Page 10

Created Date: 3/19/2009 6:27:26 PM. Mar 22, 2009  In need of the user manual for Kenwood Oscilloscope CS-4025 user manual and/or the schematic. Regards, Hello! Do you have the maintenence schematic for Kenwood Oscilloscope CS-4025? Thanks Paulo Advertisement 14th November 2007, 10:56 #17. View Forum Posts Private Message View Blog Entries. The 4025 from Kenwood, part of the CS - 4000 Series, is a Benchtop oscilloscope with a maximum bandwidth of 20 Mhz. The oscilloscope features 2 analog channels, 0 digital channels and runs on a Embedded Operating System. On Sale Kenwood CS-4125A. We take purchase orders and can provide special school, government and corporate pricing. Kenwood CS-4125A 2-Channel Oscilloscope Kenwood CS-4125A TEquipment. Kenwood service and user manuals starting with 'CS-' 1 2. KENWOOD CS1010 Service Manual. This manual has 73 pages, file size: 13.28 MB. It's available in languages: English.

10

ADJUSTMENT

To obtain the best performance, periodically calibrate the
unit. Sometimes, only one mode need be calibrated, while
at other times, all modes should be calibrated. When one
mode is calibrated, it must be noted that the other modes
may be affected. When calibrating all modes, perform the
calibration in the specified sequence.

The following calibration required an accurate measuring in-
strument and an insulated adjusting flat blade screwdriver.
If they are not available, contact your dealer. For optimum
adjustment, turn the power on and warm up the scope suffi-
ciently (more than 30 minutes) before starting.

Before calibrating the scope, check the power supply
voltage.

TEST EQUIPMENT REQUIRED
The following instrument or their equivalent should be used
for making adjustment.

Test Minimum
Equipment Model Specification
. . . Impedance: More than
Dlg'ta' Multi- 11'7“ 10 M9, Measuring range:
Meter IKENWOOD) 0.01 V to 199 V
: 1 H t
Sine-Wave 651 B :Igfifigfzonsotanzt foltage
Generator (YHPl over tuning range
Frequency: 50 kHz to
Sine-Wave 56-503 100 MHz, Output impedan-
Genarator (Tektronix) ce: 50 fl, constant voltage
over tuning range
Output signal: 1 kHz,
Amplitude: 10 mVp-p to
Square- P 506 10 Vp-p, Accuracy: within
Wave (T E' . l i 1%, Rise time: 35ns or
Generator e tronix less
100 kHz, Rise time: 1 ns or
less
43433
0 Meter (YHP) -
Color
Pattern CG'91 1A _
Generator (KENWOODl
Sensitivity: more than 5 mV
- 475A .
Oscilloscope (Tektronix) Egggugggywrlfiszponse. More
Time-Marker TG-501 Time mark: 0.5 s to 0.1ps
Generator (Tektronix) repetive waveform
High-Voltage .
Probe — Input Impedance. 1000 MIT
- - _ Impedance: 50 o
TerITllnatIOI‘l Accuracy: within 3%
~ . _ 3 watts type impedance:
Termination 50 fl
Attenuator - — 20 dB attenuation (50 m
Table 1

PREPARATION FOR ADJUSTMENT

Control Settings

The control settings listed below must be used for each
adjustment procedure.

Exceptions to these settings will be noted as they occur.
After completing a adjustment, return the controls to the fol—
lowing settings.

ADJUSTMENT

NAME OF KNOBS POSITION
MODE AUTO
COUPLING AC
SOURCE VERT
VERT MODE CH1

INV OFF _-_
SLOPE .I. +
TRIGGER LEVEL 12 o'clock
CH1/CH2 POSITION 12 o'clock
CH1/CH2 ATT 10 mV/DIV

CH1/CH2 VARIABLE Fully clockwise
CH1/CH2 AC-GND-DC DC

SWEEP POSITION 12 o'clock
SWEEP VARIABLE Fully clockwise

SWEEP ATT 0.2 ms
x10 MAG OFF _-_
FOCUS Optimum position
INTENSITY 12 o’clock after ad-
justment
Table 2

Adjustment
It
om VR (TC) P.c.B. Procedure

Supply voltage VR1 X68-1570 Adjust VR1 to give—8,00 V at pin P1-6.

Intensity VR2 X68-1570 MODE: X-Y
With a spot on the screen, turn the INTENSITY knob to a 9 o'clock posi~
tion and then adjust VR2 until the spot disappears.

CRT center VRIOZ X65-1380 CH2: 50 mV/DIV. GND
VvMODE: CH2
Pull and push INV to find a position of POSITION knob where the lumines-
cent line does not vary.
Leaving POSITION unchanged, adjust to the center of the screen using
VR102.

Y operating point VR104 X65-1380 Position after CRT center adjustment
Adjust VR104 to give + 57.0 V at pin P64.

X operating point VR301 X65—1380 MODE: X-Y
X—POSITION: Center
Adjust VR301 to give +68.0 V at pin P7—1.

ASTIG/FOCUS VR3 X68-157O MODE: X-Y
INTENSITY: 12 o'clock
FOCUS: 12 o’clock
CH1,2: 50 mV/DIV, GND
Adjust the spot ”.' to the best position with the ASTIG knob and VR3.
‘Best position = smallest spot

CH1 Step VR2 X65-1380 CH1: GND

ATT Balance Adjust VR2 so that the luminescent line does not vary when switching
between 1 mV/DIV and 2 mV/DIV.

Oscilloscope Instruction Manual

CHI MAG Balance VR3 X65-1380 CH1: GND
Adjust VR3 so that the luminescent line does not vary when switching
between 2 mV/DIV and 5 mV/DIV.

CH2 Step VRSZ X65-1380 V-MODE: CH2

ATT Balance CH2: GND
Adjust VR52 so that the luminescent line does not very when switching
between 1 mV/DIV and 2 mV/DIV.

CH2 MAG Balance VR53 X65-138O V-MODE: CH2
CH2: GND
Adjust VR53 so that the luminescent line does not vary when switching
between 2 mV/DIV and 5 mV/DIV.

ADD Position VR101 X65-1380 CH'I,2: GND
Bring the luminescent line to the center for both CH1 and CH2. Switch
V-MODE to ADD and adjust VR101 so that the luminescent line comes to
the center.

CH1 waveform VR5 X65-1380 Apply a 1 kHz square wave to CH1 INPUT (with the amplitude extending

Kenwood Cs-4025 Oscilloscope Manual

shaping VR4 over 6 div.) Adjust so that CH1 ATl' waveform is flat for both 10 mV/DIV

Dso138 Oscilloscope Manual

10 mV and 1 mV/DIV ranges.

1 mV 10 mV/DIV: VHS

1 mV/DIV: VR4

Kenwood Cs-4025 Oscilloscope Service Manual

Cs-4025

I I

Oscilloscope Manual Pdf

*l+*l